I-GDOT-80A yoMvavanyi we-oyile yokugquma i-oyile ihlaziywe1105
●Nceda ufunde incwadi yokusebenza ngononophelo phambi kokuba usebenze.
●Nceda ujonge ukuba umvavanyi uqhagamshelwe kakuhle emhlabeni phambi kovavanyo.
●Akuvumelekanga ukuhambisa okanye ukuphakamisa ikhava yokuvavanya kwinkqubo yokuvavanya ukuphepha ukulimala ngombane ophezulu.Umbane kufuneka ucinywe phambi kokuba kubuyiselwe ioyile yesampulu.
●Phatha ngononophelo xa ukhulula okanye uvala isiciko sovavanyo lombane ophezulu!
●Ukuba umvavanyi usebenza ngokungaqhelekanga emva kokuba i-oyile yokugquma ingasebenzi, nceda ucime umvavanyi kangangemizuzwana eyi-10, emva koko uqalise kwakhona.
●Emva kokuba iphepha lokuprinta liphelile, nceda ujonge inxalenye yenkcazo yomshicileli (okanye isihlomelo semanuwali) ukutshintsha iphepha lokuprinta ukunqanda umonakalo wentloko yomshicileli.
●Gcina umvavanyi kude nokufuma, uthuli kunye nezinye izinto ezonakalisayo, kwaye uyigcine kude nemithombo yobushushu obuphezulu.
●Phatha ngononophelo kwizothutho.Musa ukubeka icala phantsi.
●Incwadana inokuhlaziya ngokufanelekileyo ngaphandle kwesaziso esinye kwangaphambili.Ukuba nawuphi na umbuzo, nceda uqhagamshelane nathi.
Ixesha lewaranti yolu chungechunge ngunyaka omnye ukususela kumhla wokuthunyelwa.Nceda ubhekisele kwi-invoyisi yakho okanye amaxwebhu okuthumela ukuze ubone imihla efanelekileyo yewaranti.I-HVHIPOT iqinisekisa umthengi wokuqala ukuba le mveliso ayinaziphene kwizinto eziphathekayo kunye nokusebenza phantsi kokusetyenziswa kwesiqhelo.Kulo lonke ixesha lewaranti, ukubonelela ukuba ezo ziphene azinqunywanga yi-HVHIPOT ukuba ibangelwa ukusetyenziswa kakubi, ukusetyenziswa kakubi, ukuguqulwa, ukufakela okungafanelekanga, ukungahoywa okanye imeko embi yendalo, i-HVHIPOT inqunyelwe kuphela ukulungiswa okanye ukutshintshwa kwesi sixhobo ngexesha lewaranti.
| Isixhobo se-GDOT-80C | 1 pc |
| Ikomityi yeoli(250ml) | iipcs ezi-1 |
| Ikhodi yamandla | 1 pc |
| I-fuse esecaleni | iipcs ezi-2 |
| Ukubetha intonga | iipcs ezi-2 |
| Igeyiji eqhelekileyo(25mm) | 1 pc |
| Iphepha lokuprinta | Imiqulu emi-2 |
| Tweezer | 1 pc |
| Isikhokelo somsebenzisi | 1 pc |
| Ingxelo yovavanyo lwasefektri | 1 pc |
HV Hipot Electric Co., Ltd. uye ngokungqongqo nangokucokisekileyo proofread manual, kodwa asinakuqinisekisa ukuba akukho ziphoso kunye nezinto ezishiyiweyo ngokupheleleyo kwincwadana.
I-HV Hipot Electric Co., Ltd. izimisele ukwenza uphuculo oluqhubekayo kwimisebenzi yemveliso, kunye nokuphucula umgangatho wenkonzo, ngoko ke inkampani ihlala inelungelo lokutshintsha naziphi na iimveliso kunye neenkqubo zesoftware ezichazwe kule ncwadana kunye nomxholo wale ncwadana isaziso.
Ukugquma kwangaphakathi kwezixhobo zombane ezininzi kwiinkqubo zamandla, iinkqubo zikaloliwe, izityalo ezinkulu zepetrochemical kunye namashishini, uninzi lwazo lwamkelwa uhlobo lokugquma oluzaliswe yioyile, ke ngoko, uvavanyo lwamandla e-dielectric oyile luqhelekile kwaye luyimfuneko.Ukuhlangabezana neemfuno zemarike, siye saphuhlisa kwaye savelisa uluhlu lwabavavanyi bamandla e-oyile ye-dielectric ekhuselayo ngokwemigangatho yesizwe ye-GB/T507-2002, umgangatho woshishino DL429.9-91 kunye nomgangatho wamva nje woMbane woMbane woMbane woMbane weDL/T846.7 -2004 ngokwethu.Esi sixhobo, ngokusebenzisa i-microcomputer ye-single-chip njengeyona nto ingundoqo, yaqonda ukusebenza okuzenzekelayo, umlinganiselo ochanekileyo ophezulu, uphucula kakhulu ukusebenza kakuhle kunye nokunciphisa ubuninzi bomsebenzi wabasebenzi.Ngaphezu koko, incinci ngobukhulu kwaye ikulungele ukuyiphatha.
●Nge-microprocessor, zalisekisa ngokuzenzekelayo uvavanyo lwe-voltage yokumelana nokujikeleza kweoli kunye noluhlu lwe-0 ~ 80KV (kubandakanywa ukunyusa, ukugcinwa, ukuxuba, ukuma, ukubala, ukuprinta kunye neminye imisebenzi).
●Isibonisi esikhulu sesikrini se-LCD.
●Ukusebenza okulula.Umatshini uya kugqibezela ngokuzenzekelayo uvavanyo lokumelana nombane kwikomityi enye yeoli yesampulu emva kokusetwa lula ngumqhubi.Ixabiso le-voltage yokuqhawula amaxesha angama-1 ~ 6 kunye namaxesha okujikeleza aya kugcinwa ngokuzenzekelayo.Emva kovavanyo, umshicileli we-thermal uya kuprinta ixabiso ngalinye le-voltage yokuqhawula kunye nexabiso eliphakathi.
●Ukugcinwa kwamandla phantsi.Ingagcina iziphumo zovavanyo ezili-100 kwaye ubonise ubushushu obukhoyo kunye nokufuma.
●Yamkela i-microcomputer ye-single-chip ukonyusa amandla ombane ngesantya esingaguqukiyo.I-voltage frequency ichanekile kwi-50HZ, qinisekisa ukuba yonke inkqubo kulula ukuyilawula.
●Nge-over-voltage, over-current and limit limit protection ukuqinisekisa ukhuseleko lwabaqhubi.
●Ngomsebenzi wokubonisa ubushushu obulinganisiweyo kunye newotshi yenkqubo.
●Nxibelelana nekhompyuter ngojongano oluqhelekileyo lweRS232.
| Ukunikezwa amandla | AC220V±10%, 50Hz |
| Imveliso yombane | 0~80kV(Okukhethwayo) |
| Umthamo | 1.5kVA |
| Amandla | 200W |
| Isantya sokunyuka kwamandla ombane | 2.0~3.5kV/s (Inokulungiswa) |
| I-Voltageukulinganisaukuchaneka | ±3% |
| Ukuguqulwa kwamaza | ≤3% |
| Ikhefu lokunyusa | 5min (Inokulungiswa) |
| Ixesha lokuma | 15min (Inokulungiswa) |
| Amaxesha okunyusa | 1~6 (Ekhethiweyo) |
| Ukusebenzaokusingqongileyo | Tubushushu: 0℃-45°C |
| Hubude:Max.Ubuntu bezizalwane75% | |
| Ubungakanani | 465x385x425mm |

① Umshicileli we-Thermal--ukuprinta iziphumo zovavanyo;
② LCD--bonisa imenyu, ngokukhawuleza kunye neziphumo zovavanyo;
③ Izitshixo zokusebenza:
Cofa iqhosha elithi "◄" ukwandisa ixabiso lokumisela;
Cofa "►" iqhosha lokunciphisa ixabiso lokusetha;
Khetha--yokukhetha imisebenzi (inqaku elikhethiweyo likumboniso ogciniweyo);
Qinisekisa--ukwenza imisebenzi;
Emva--yokuphuma kujongano lokusebenza;
④ Ukutshintsha kwamandla kunye nesalathisi
1. Ukulungiselela phambi kovavanyo
1.1 Qhagamshela i-terminal yokumisa (kwicala lasekunene lesixhobo) kucingo olusezantsi ngokuqinileyo phambi kokusebenzisa izixhobo.
1.2 Isampuli yeoli yokukhupha ngokwemigangatho efanelekileyo.Lungisa umgama we-electrode ngaphakathi kwekomityi yeoli ngokwegeyiji eqhelekileyo.Coca indebe ngokweemfuno ezifanelekileyo.Galela isampuli yeoli kwikomityi kwaye uvale i-cap.
1.3 Ukutshintshela umbane we-AC220V emva kokuba ezi zinto zingentla ziqinisekisiwe, zilungele uvavanyo.
2. Uvavanyo
2.1 Cofa iswitshi yamandla uze ungene kujongano olulandelayo:

2.2 Ukusetwa kweParamitha yeSistim

Cofa iqhosha elithi "Qinisekisa" kwaye ungene kujongano olulandelayo:

Ukusetwa kokunyusa: abasebenzisi banokukhetha ngokwemfuno yokwenyani.

Cofa iqhosha elithi "Buyela" ukuphuma kolu jongano emva kokuba useto lwenziwe.
2.3 Uvavanyo
Cofa iqhosha elithi "Khetha" ukukhetha imenyu ethi "Qala uvavanyo" kwaye ucinezele iqhosha elithi "Qinisekisa" ukungena kujongano olulandelayo:



Ukuqhubeka novavanyo olulandelayo ngokukhawuleza nje ukuba uvavanyo lokuqala lugqityiwe de kugqitywe i-frequency boosting frequency.Ekugqibeleni, isiphumo siyaboniswa kwaye siprintwe ngolu hlobo lulandelayo:

2.4 Ukujongwa kwedatha kunye noshicilelo:
Cofa iqhosha elithi "Khetha" ukukhetha imenyu "yokuJonga nokuprinta" kwaye ucinezele iqhosha elithi "Qinisekisa" ukungena kujongano olulandelayo:

Khetha u-"Page Up" okanye "Page Down" kwaye ukhethe iirekhodi eziza kuprintwa kwaye ukhethe "Shicilela".
●Ukukhethwa kwesampulu yeoli kunye nokubekwa komgama we-electrode kuya kuhlangabezana nemigangatho efanelekileyo yelizwe kunye neyeshishini.
●Abaqhubi okanye abanye abasebenzi bathintelwe ngokungqongqo ukuba bachukumise iqokobhe emva kokuba umbane uvuliwe ukunqanda iingozi.
●Umbane uya kunqunyulwa ngokukhawuleza ukuba kukho nasiphi na isiganeko esingaqhelekanga esifunyenwe ngexesha lokusebenza.
●Esi sixhobo asiyi kubonakaliswa kwindawo efumileyo.
●Gcina ikomityi yeoli kunye nee-electrode zicocekile.Gcwalisa iikomityi eneoli entsha yokukhusela xa ingasebenzi.Jonga umgama we-electrode kwaye ujonge ukuxinana phakathi kwencam ye-electrode kunye ne-electrode bar screw thread ngaphambi kokuba ikomityi isetyenziswe kwakhona.
1. Indlela yokucoca indebe ye-oyile
1.1 Sula umphezulu we-electrode kunye neentsimbi kwakhona kwaye kwakhona ngelaphu lesilika elicocekileyo.
1.2 Lungisa umgama we-electrode ngegeyiji eqhelekileyo
1.3 Sebenzisa i-petroleum ether (ezinye izinyibilikisi ze-organic azivumelekanga) ukucoca kathathu.Ngalo lonke ixesha kuya kulandelwa ezi nkqubo zilandelayo:
① Galela i-ether yepetroleum kwikomityi yeoli de ikomityi ibe yi-1/4~1/3 igcwele.
② Gquma udini lwekomityi ngeqhekeza leglasi elicocwe ngepetroleum ether.Gubha indebe ngokulinganayo ngomzuzu omnye ngamandla athile.
③ Galela kude i-ether yepetroleum kwaye womise ikomityi ngesivuthela imizuzu emi-2-3.
1.4 Sebenzisa isampulu yeoli ukuba ivavanywe ukucoca ikomityi kangange 1 ~ 3 amaxesha.
① Galela i-ether yepetroleum kwikomityi yeoli de ikomityi ibe yi-1/4~1/3 igcwele.
② Gquma udini lwekomityi ngeqhekeza leglasi elicocwe ngepetroleum ether.Gubha indebe ngokulinganayo ngomzuzu omnye ngamandla athile.
③ Galela kude isampuli yeoli ekhohlo kwaye uvavanyo luqalise.
2. Indlela yokucoca intonga yokuvuselela
2.1 Sula intonga yokuvuselela kwakhona kwaye kwakhona ngelaphu lesilika elicocekileyo kude kube ngamasuntswana amahle angafumaneki kumphezulu wawo.Akuvumelekanga ukubamba imiphezulu ngezandla.
2.2 Sebenzisa izinyanzelo ukucinezela intonga;zifake kwi-ether yepetroleum kwaye uhlambe.
2.3 Sebenzisa iiforcep ukucinezela intonga uze uyomise nge blower.
2.4 Sebenzisa izinyanzelo ukucinezela intonga;zifake kwisampulu yeoli kwaye uhlambe.
3. Ukugcinwa kwe-Oil Cup
Indlela yoku-1 Gcwalisa indebe ngeoli ye-insulating efanelekileyo emva kokuba uvavanyo lugqityiwe kwaye lubeke luzinzile.
Indlela yesi-2 Coca kwaye womise ikomityi phantsi kwezi nkqubo zingasentla uze uyibeke kwi-vacuum dryer.
Qaphela: Ikomityi yeoli kunye nentonga yokugubha iya kuhlanjululwa phantsi kweenkqubo ezingentla emva kovavanyo lokuqala kunye neemvavanyo ngeoli embi.
4. UCwangciso lweFault oluqhelekileyo
4.1 Umbane ucinyiwe, akukho sibonisi kwisikrini
① Jonga ukuba iplagi yamandla ifakwe nkqi okanye hayi.
② Jonga ukuba i-fuse engaphakathi kwindawo yombane ikwimeko entle okanye ayikho.
③ Jonga umbane wesokethi.
4.2 Ikomityi ye-oyile ngaphandle kokuphazamiseka
① Khangela ukufakwa kweziqhagamshelo kwibhodi yesekethe.
② Jonga uqhagamshelwano lweswitshi ephezulu yombane kwikhava yetyala.
③ Jonga ukusebenza koqhagamshelwano lombane ophezulu.
④ Jonga ikhefu lomgca we-high-voltage.
4.3 Umahluko wokuboniswa akwanelanga
Lungisa i-potentiometer kwibhodi yesekethe.
4.4 Ukusilela kweprinta
① Jonga iplagi yamandla yomshicileli.
② Jonga iplagi yomgca wedatha yomshicileli.