I-HV HIPOT GDC-9560B Inkqubo yamandla yokuSula iOyile yegesi yeChromatography Analyzer
Ukufakela kunye nokususwa kwekholamu yechromatographic:
1. Ukufakwa kunye nokususwa kwekholomu yechromatographic kufuneka kuqhutywe kwiqondo lokushisa.
2. Iinqaba ezipakishiweyo zinezitywina ze-ferrule kunye ne-gasket seals.Kukho iintlobo ezintathu zeeferrules: iiferrules zetsimbi, iiferrules zeplastiki kunye negraphite ferrules, ekungelula ukuba ziqiniswe kakhulu xa ufaka.Iitywina zohlobo lwe-gasket zifuna iigaskets ezintsha ngexesha ngalinye kufakwe ikholamu.
3. Ingaba iziphelo ezibini zekholamu yechromatographic zixhunyiwe ngoboya beglasi.Thintela uboya beglasi kunye nokupakishwa ekubeni kungavuthelwa kwisixhobo somoya yirhasi yokuthwala.
4. Ubude bokufakwa kwekholamu ye-capillary kunye nokufakwa kuxhomekeke kwincwadana yomyalelo wesixhobo.Ubume bamagumbi ahlukeneyo e-chromatographic evaporation ahlukile, kwaye ubude bokufaka buhluke.Kufuneka kuqatshelwe ukuba ikholomu ye-capillary isetyenziswe kunye ne-splitless flow, i-interface phakathi kwegumbi le-vaporization kunye nekholomu epakishweyo akufanele ibe neeprobes ezininzi ezixhunyiwe kwigumbi lokuphunga, kwaye ikholamu ye-capillary kufuneka ibe ngaphaya kwe-cap.
Iimpembelelo zomlinganiselo we-hydrogen ukuya emoyeni kwi-FID detector:
Kwi-chromatograph yegesi, umlinganiselo we-hydrogen ukuya emoyeni kufuneka ube ngu-1:10.Ukuba umlinganiselo uphezulu kakhulu, uvakalelo lwe-hydrogen flame detector luya kuhla ngokukhawuleza.Jonga amazinga e-hydrogen kunye nomoya wokuhamba.Xa igesi kwi-hydrogen kunye nomoya ingonelanga, iya kwenza "i-bang" ukuba ivuthe, kwaye emva koko ikhuphe umlilo, ngokuqhelekileyo xa ukhanyisa isitofu sombane, siyacinywa, kwaye siphinde sikhanyise, kwaye emva koko. ihydrogen ayonelanga.
Ukuthintela njani inaliti yokutofa ukuba ingagobi:
Uninzi lwabaqalayo kwichromatography bahlala begoba inaliti kunye nesiqu kuba:
1. I-injection port ye-chromatograph yegesi ikrwelwe ngokuqinileyo, kwaye ikrwelwe ngokuqinileyo kwiqondo lokushisa.Xa ubushushu begumbi le-vaporization liqhubeka nokunyuka, i-silicone gasket iya kwanda kwaye iqinise.Ngeli xesha, kunzima ukungena ngesirinji.
2. Inaliti ibambelele kwinxalenye yentsimbi ye-injection port ukuba indawo ayilula ukuyifumana.
3. Intonga yesirinji igobile ngenxa yokuba amandla anamandla kakhulu xa utofa isampuli.Kukho i-sampler rack yokungenisa ichromatographic.Akunakwenzeka ukugoba intonga yesirinji yesiqhelo xa usebenzisa i-sampler rack ukufaka iisampuli.
4. Ekubeni udonga lwangaphakathi lwesirinji lungcolisekile, tyhala i-shaft yenaliti ukugoba ngexesha lokutofa.
5. I-chromatograph yegesi kufuneka izinzile ngexesha lokutofa.Ukuba ungxamile, uya kuyigoba isirinji.Ukuba uqhelene nalo msebenzi, uya kugqiba ngokukhawuleza.